Fiber Optic Quick Coupling High Speed Handpiece

Extract the bur from the handpiece, use the soft brush to clean the dirt, and wash with disinfection of alcohol cotton.
Lubricate the handpiece, pour two drops into the air tube one time.
Wrap it with a sterilization package and mark.

Model No

F3-K6

Features

Kavo type, with Quick Coupling

Chuck Type

Push Button

Working Air Pressure

0.20-0.25MPa(0.22MPa preferred)

Rotation speed

320,000-400,000rpm

Bur Applicable

Φ1.595-1.600mm

Spray

Triple Water Spray

Holes

4, 6

Luminance

16000 LUX -20000 LUX

Temperature

3500 K -4000 K

Details about Fiber Optic Quick Coupling High Speed Handpiece

After each treatment, Sterilizing handpiece in an autoclave according to the following steps:

1.Remove the bur and brush clean handpiece, clean outside of handpiece with an alcohol solution.
2.Lubricate handpiece, each time drop two drops inside of air inlet pipe.
3.Place into autoclaving pouch and mark.
4.Autoclave the handpiece for 12 minutes at 135℃.